Graham Field NEB-U-TYKE JB0112-164 User Manual Page 10

  • Download
  • Add to my manuals
  • Print
  • Page
    / 32
  • Table of contents
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 9
John Bunn Neb-u-Tyke RR Train Nebulizer Compressor 10
Cleaning the Nebulizer Cup
All parts of the disposable nebulizer, except tubing, should
be thoroughly cleaned after each use. If your physician
species a specic cleaning procedure, follow that proce-
dure.
WARNING: To prevent possible risk of infection from con-
taminated medication, clean the nebulizer after each treat-
ment.
1. After each use, disassemble the nebulizer cup, T adapt-
er and mouthpiece completely, wash all parts in warm
soapy water, and rinse thoroughly. Allow to air dry.
Store in a clean, dry location.
2. Once daily, after washing as above, soak nebulizer com-
ponents in a fresh solution consisting of one part warm
water and one part white vinegar for 30 minutes fol-
lowed by a warm water rinse. Allow to air dry. Store in a
clean, dry location.
Replacing the Filter
Note: Change the filter once every 30 days or when the
filter changes color. Additional filters can be purchased at
www.grahamfield.com or from your Graham-Field authorized
distributor.
1. Ensure that ON/OFF switch is in the OFF position. Open
the lter cap located next to the ON/OFF switch.
2. Remove the dirty lter with a small, pointed object. Dis-
card the lter.
3. Replace with a clean lter.
4. Replace the lter cap.
s
Caution: Do not use cotton or other materials as a filter.
s
Caution: Do not wash or clean the filter.
s
Caution: Do not reuse the filter. Do not operate com-
pressor without a filter. Compressor damage can result.
s
Caution: The disposable nebulizer kit is for single pa-
tient use only.
Page view 9
1 2 ... 5 6 7 8 9 10 11 12 13 14 15 ... 31 32

Comments to this Manuals

No comments